Impact of polyurethane on variations in tracheal cuff pressure in critically ill patients: a prospective
Saad Nseir1, Farid Zerimech, Julien De Jonckheere
1Intensive Care Unit, Calmette Hospital, University Hospital of Lille, boulevard du Pr Leclercq, 59037, Lille Cedex, France. s-nseir@chru-lille.fr
Objective:
To determine the impact of polyurethane (PU) on variations in cuff pressure (P (cuff)) in intubated critically ill patients.
Methods:
Prospective observational before-after study performed in a ten-bed ICU. Cuff pressure was continuously recorded for 24 h in 76 intubated patients, including 26 with polyvinyl chloride (PVC), 22 with cylindrical polyurethane (CPU), and 28 with tapered polyurethane (TPU)-cuffed tracheal tubes. P (cuff) was manually adjusted every 8 h by nurses and was maintained around 25 cmH(2)O. Time spent with cuff underinflation and overinflation was continuously measured. In addition, pepsin, a proxy for microaspiration of gastric contents, was quantitatively measured in tracheal secretions at the end of recording period.
Results:
A total of 1,824 h of continuous recording of cuff pressure was analyzed. Patient characteristics were similar in the three groups. No significant difference was found in percentage of time spent with underinflation (mean +/- SD, 26 +/- 22, 28 +/- 12, 30 +/- 13% in PVC, CPU, and TPU groups, respectively) and overinflation [median (IQR), 7 (2-14), 6 (3-14), 11% (5-20)] among the three groups. However, a significant difference was found in the coefficient of variation of P (cuff) (mean +/- SD, 82 +/- 48, 92 +/- 47, 135 +/- 67, p = 0.002). While the coefficient of P (cuff) variation was significantly (p < 0.017) higher in the TPU compared to CPU and PVC groups, no significant difference was found between the CPU and PVC groups. The pepsin level was significantly different among the three groups (408 +/- 282, 217 +/- 159, 178 +/- 126 ng/ml; p < 0.001). In fact, the pepsin level was significantly lower in the CPU and TPU groups compared with the PVC group.
Conclusion:
PU does not impact variations in P (cuff) in critically ill patients.
Related Concept Videos
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.