Related Experiment Video
Updated: Jun 13, 2026

A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
Performance characteristics of five cardiac Troponin I assays
Sonia L La'ulu1, William L Roberts
1ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, UT, United States.
Background:
The aim of this study was to evaluate the performance characteristics of cardiac Troponin I (cTnI) assays on the Access 2, Advia Centaur, Architect i1000(SR), Architect i2000(SR), and Vitros ECi analyzers.
Methods:
To determine the 99th percentile, plasma and serum samples were drawn from 400 apparently healthy subjects. Functional sensitivity was evaluated by testing 10 panels of serum pools in duplicate for 10days using 2 reagent lots. Imprecision was also assessed using commercial quality control materials. For method comparison, 100 plasma samples were tested by all methods with the Architect i2000(SR) assay as the comparison method.
Results:
The 99th percentile ranged from 0.013 to 0.020microg/l. Calculated CVs at the 99th percentile were 23-52%. cTnI concentrations at a 10% CV ranged from 0.039 to 0.101microg/l. Total CVs determined using quality control materials were 2.4-10.0%. Method comparison results by Deming regression had slopes of 1.03-1.51. The mean bias for cTnI samples between 0.01 and 1.0microg/l ranged from 0 to 0.2microg/l.
Conclusions:
No method demonstrated a concentration with a 10% CV that was less than the 99th percentile for healthy adults. All methods compare well to the Architect i2000(SR) assay with the Advia Centaur having the highest positive bias.
More Related Videos
07:40Impact of High-intensity Interval Exercise and Moderate-Intensity Continuous Exercise on the Cardiac Troponin T Level at an Early Stage of Training
Published on: October 10, 2019
09:23Using Extraordinary Optical Transmission to Quantify Cardiac Biomarkers in Human Serum
Published on: December 13, 2017
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Acute Coronary Syndrome III: Diagnostic Studies
Acute Coronary Syndrome I: Introduction
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...
Specialized Characteristics of Cardiac Muscles
Cardiac muscle cells are smaller than skeletal muscles, averaging 10–20 mm in diameter and 50–100 mm in length. However, they have large energy demands for continuous contraction and relaxation. This energy is almost exclusively derived from aerobic metabolism of energy reserves in...