How can we safely use ultrasonic laparoscopic coagulating shears?
Jae Sik Kim1, Ryohei Hattori, Tokunori Yamamoto
1Department of Urology, Nagoya University Graduate School of Medicine, Nagoya, Japan.
Summary
Ultrasonic laparoscopic coagulation shears (LCS) can significantly heat the tissue pad during activation. Limiting activation time to within 5 seconds is a safe practice to protect surrounding tissues.
Area of Science:
- Laparoscopic surgery
- Surgical instrumentation
- Biomedical engineering
Background:
- Ultrasonic laparoscopic coagulation shears (LCS) are widely used in minimally invasive surgery.
- Understanding thermal spread from these devices is crucial for patient safety.
- The tissue pad, distinct from the active blade, can also generate heat during LCS activation.
Purpose of the Study:
- To quantify temperature changes in the tissue pad of ultrasonic LCS during activation.
- To evaluate histological alterations in surrounding tissues near the tissue pad.
- To compare thermal profiles of different LCS models.
Main Methods:
- Three ultrasonic LCS devices (SonoSurg, AutoSonix, Harmonic Scalpel) were tested at varying power levels and durations (5-20 seconds).
- Tissue pad temperatures were measured during extracorporeal chicken tissue cutting.
- Histological changes around the tissue pad were assessed in a porcine laparoscopic model using the SonoSurg device.
Main Results:
- Tissue pad temperatures increased with activation duration and power, reaching up to 100°C.
- Significant histological changes in surrounding tissues were observed after 10 seconds of activation.
- All tested LCS models showed temperature increases, with variations based on device and settings.
Conclusions:
- Surgeons must be aware of and account for heat generated by the tissue pad of ultrasonic LCS.
- Careful consideration of surrounding tissues and organs near both the active blade and tissue pad is essential.
- Utilizing short activation times (≤5 seconds) is a recommended safe practice, irrespective of LCS type or power setting.