Effects of imposed resistance on tidal volume with 5 neonatal nasal continuous positive airway pressure systems
Shannon E Cook1, Katherine L Fedor, Robert L Chatburn
1Respiratory Institute, The Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA.
Background:
Neonates with respiratory distress syndrome are often treated with nasal continuous positive airway pressure (CPAP). Nasal CPAP methods include electronic feedback control, underwater seal, flow opposition, and flow opposition with fluidic flow reversal on expiration. Few studies have compared those modes, and the results have been contradictory.
Methods:
We compared the effect of resistive load on simulated tidal volume (V(T)) with 5 neonatal nasal CPAP systems: Fisher and Paykel nasal CPAP tubing with Maquet Servo-i ventilator in NIV CPAP mode; Cardinal Health AirLife nasal CPAP system; Fisher and Paykel nasal CPAP tubing with water-seal pressure generator; AirLife infant nasal CPAP generator kit; and Hamilton Medical Arabella fluidic nasal CPAP generator. The lung simulator settings were: compliance 0.5 mL/cm H2O, resistance 125 H2O/L/s, sinusoidal patient-effort range 6.5-26 cm H2O, rise 25%, hold 0%, release 25%, respiratory rate 65 breaths/min. We compared the mean values from 10 breaths.
Results:
The mean inspiratory pressure drop and V(T) difference (compared to the simulator alone, unloaded) increased with V(T), respectively, from 0.32 cm H2O to 1.73 cm H2O, and from -0.04 mL to -0.40 mL. Flow opposition had the smallest pressure drop (from 0.10 cm H2O to 0.64 cm H2O, P < .001). At V(T) of < or = 6 mL, the bubble nasal CPAP's pressure drop was largest (P < .001), whereas at V(T) of > or = 9 mL the electronic nasal CPAP's pressure drop was largest (P < .001). All systems except the ventilator did not have an average end-expiratory pressure of the targeted 5 cm H2O.
Conclusions:
The differences in these nasal CPAP systems correlate with the differences in unassisted V(T) due to loading effects. The ventilator imposed the least load, and the AirLife nasal CPAP system imposed the most.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...