Development of a new control device for stabilizing blood level in reservoir during extracorporeal circulation
Naoki Momose1, Rie Yamakoshi, Ryo Kokubo
1Department of Medical Engineering, Saitama Medical Center, Jichi Medical School, Omiya, Japan.
Abstract:
We developed a simple device that stabilizes the blood level in the reservoir of the extracorporeal circulation open circuit system by measuring the hydrostatic pressure of the reservoir to control the flow rate of the arterial pump. When the flow rate of the venous return decreases, the rotation speed of the arterial pump is automatically slowed down. Consequently, the blood level in the reservoir is stabilized quickly between two arbitrarily set levels and never falls below the pre-set low level. We conducted a basic experiment to verify the operation of the device, using a mock circuit with water. Commercially available pumps and reservoir were used without modification. The results confirmed that the control method effectively regulates the reservoir liquid level and is highly reliable. The device possibly also functions as a safety device.