Interference of cardiac pacemaker and implantable cardioverter-defibrillator activity during electronic dental device
Jason J Roedig1, Jignesh Shah, Claude Samy Elayi
1Department of Oral Health Practice, College of Dentistry, University of Kentucky, Lexington, KY 40536-0084, USA.
Background:
The authors conducted a study to determine if electromagnetic interference of cardiac pacemaker and implantable cardioverter-defibrillator (ICD) activity occurs during the operation of electronic dental devices.
Methods:
The authors tested nine electronic dental devices in vitro to assess their ability to interfere with the function of two pacemakers and two ICDs as determined by electrocardiographic telemetry.
Results:
The pacing activity of both pacemakers and the dual-chamber ICD were inhibited during operation of the battery-operated composite curing light at between 2 and 10 centimeters from the generator or leads. The use of the ultrasonic scaler interfered with the pacing activity of the dual-chamber pacemaker at between 17 and 23 cm from the generator or leads, the single-chamber pacemaker at 15 cm from the generator or leads and both ICDs at 7 cm from the leads. The operation of the ultrasonic cleaning system interfered with the activity of the dual-chamber pacemaker at between 15 and 23 cm from the generator or leads, and of the single-chamber pacemaker at 12 cm. Operation of the electric toothbrush, electrosurgical unit, electric pulp tester, high- and low-speed handpieces, and an amalgamator did not alter pacing function.
Conclusion:
Select electronic dental devices interfere with pacemakers' and ICDs' sensing and pacing activity in vitro.
Clinical Implications:
Use of the ultrasonic scaler, ultrasonic cleaning system and battery-operated composite curing light may produce deleterious effects in patients who have pacemakers or ICDs.
Related Concept Videos
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Mechanism of Cardiac Arrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Antiarrhythmic Drugs: Class IV Agents as Calcium Channel Blockers
Verapamil, a calcium channel blocker, inhibits calcium movement across myocardial cell membranes and vascular smooth muscle. This results in the dilation of coronary and...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias VI: Management of Dysrhythmias