Related Experiment Video
Updated: Jun 13, 2026

Lateral-PLIF for Lumbar Spinal Arthrodesis: A Detailed Step-By-Step Surgical Technique
Published on: January 23, 2026
Analysis of biomechanical changes after removal of instrumentation in lumbar arthrodesis by finite element analysis
Ho-Joong Kim1, Heoung-Jae Chun, Seong-Hwan Moon
1Yonsei sarang Hospital, Seoul, Korea.
Abstract:
The purpose of this study is to investigate the change in biomechanical milieu following removal of pedicle screws in instrumented single level lumbar arthrodesis. Using a validated finite element (FE) model of the intact lumbar spine (L2-5), two scenarios of L3-4 lumbar fusion were simulated: posterolateral fusion (PLF) at L3-4 using pedicle screws (PLF with pedicle screws; WiP) and L3-4 lumbar posterolateral fusion state after removal of pedicle screws (PLF without pedicle screws; WoP). The WiP model had greater range of motion (ROM) at each adjacent segment than the WoP model. This phenomenon became pronounced at the proximal adjacent segment under flexion moment. Similarly, removal of pedicle screws (the WoP model) relieved the maximal von Mises stress at adjacent segments under 4 moments compared to the WiP model. This study demonstrated that removal of pedicle screws could decrease stiffness of fusion segments, which would reduce the disk stress of adjacent segments.