Absolute calibration of LIDAR Thomson scattering systems by rotational Raman scattering
R Scannell1, M Beurskens, M Kempenaars
1EURATOM/CCFE Fusion Association, Culham Science Centre, Abingdon OX14 3DB, United Kingdom. rory.scannell@ccfe.ac.uk
Abstract:
Absolute calibration of LIDAR Thomson scattering systems on large fusion devices may be achieved using rotational Raman scattering. The choice of calibrating gas molecule presents different options and design trade-offs and is likely to be strongly dependent on the laser wavelength selected. Raman scattering of hydrogenic molecules produces a very broad spectrum, however, with far fewer scattered photons than scattering from nitrogen or oxygen at the same gas pressure. Lower laser wavelengths have the advantage that the Raman cross section increases, sigma(Raman) proportional to 1/lambda(0)(4), but the disadvantage that the spectral width of the scattered spectrum decreases, Deltalambda(Raman) proportional to lambda(0)(2). This narrower spectrum makes measurement closer to the laser wavelength necessary. The design of the calibration technique presents a number of challenges. Some of these challenges are generic to all Thomson scattering systems. These include detecting a sufficient number of photoelectrons and designing filters that measure close to the laser wavelength while simultaneously achieving adequate blocking of the laser wavelength. An issue specific to LIDAR systems arises since the collection optics operates over a wide range of depth of field. This wide depth of field has the effect of changing the angle of light incident on the optical interference filter with plasma major radius. The angular distribution then determines the effective spectral transmission function of the interference filter and hence impacts on the accuracy of the absolute calibration. One method that can be used to increase absolute calibration accuracy is collecting both Stokes and anti-Stokes lines with optical filter transmission bands specifically designed to reduce systematic uncertainty.
More Related Videos
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...