Differences in detecting blasts between ADVIA 2120 and Beckman-Coulter LH750 hematology analyzers
S G Shelat1, W Canfield, S Shibutani
1Department of Pathology and Laboratory Medicine, Philadelphia, PA 19104, USA. shelat@mail.med.upenn.edu
International Journal of Laboratory Hematology
|May 8, 2010
Summary
The Siemens ADVIA 2120 shows superior sensitivity in detecting leukemia blast cells compared to the Beckman-Coulter LH750. However, careful manual review remains crucial due to differing specificities in blast cell detection.
Area of Science:
- Hematology
- Clinical Pathology
- Laboratory Medicine
Background:
- Complete blood count (CBC) is a routine screening tool for hematologic disorders.
- Blast cells in peripheral blood are key indicators in leukemia diagnosis and monitoring.
- Automated hematology analyzers are increasingly used for initial detection of abnormal cells.
Purpose of the Study:
- To compare the analytical performance of two hematology analyzers, Siemens ADVIA 2120 and Beckman-Coulter LH750.
- To evaluate their accuracy in detecting blast cells in peripheral blood samples.
- To assess the impact of analyzer performance on laboratory diagnostic workflows.
Main Methods:
- Comparative analytical study of two hematology analyzers.
- Analysis of 390 peripheral blood samples from a general hospital population, including oncology patients.
- Comparison of analyzer blast cell flags against manual differential counts as the reference method.
Main Results:
- Siemens ADVIA 2120 achieved 100% sensitivity for blast cell detection, significantly higher than the Beckman-Coulter LH750's 62%.
- The ADVIA 2120 had lower specificity (49%) compared to the LH750 (86%).
- Laboratory review criteria effectively managed false positives and negatives from both analyzers.
Conclusions:
- The Siemens ADVIA 2120 offers enhanced sensitivity for detecting peripheral blood blasts, crucial for leukemia screening.
- Analyzer performance characteristics, combined with stringent manual review protocols, are vital for accurate blast cell identification.
- Optimizing analyzer selection and laboratory workflows can improve the efficiency and accuracy of hematologic abnormality detection.