Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Metal artifacts in MRI from non-magnetic dental alloy and its FEM analysis
Tomohide Taniyama1, Taiji Sohmura, Takanori Etoh
1Division of Oromaxillofacial Regeneration, Course for Integrated Oral Sciences and Stomatology, Osaka University Graduate School of Dentistry, Osaka, Japan. tom-tani-dental@occn.zaq.ne.jp
Abstract:
Artifacts in MR(Magnetic Resonance) images of oral cavity produced from non-magnetic metal restorations was verified by measuring the image of index finger and a cylinder of fat test piece with a type 4 gold alloy ring using a compact MRI equipment. In the images of finger, portion around the ring disappeared. However, it was nearly restored with a cut ring. In the cylinder of fat test piece, obvious artifacts appeared when circumferential surface of the ring was placed perpendicular to RF(Radio Frequency) field of MRI equipment's excitation/detection coil. However, in other directions or with a cut ring, artifact disappeared. The cause was simulated with FEM(Finite Element Method) electromagnetic field analysis, and alternating magnetic field was shown to induce surface current on the continuous gold ring. Magnetic field produced by that current interfered with the field from excitation coil. This demonstrated the characteristics and cause of artifacts by non-magnetic dental metals.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...