Related Experiment Video
Updated: Jun 13, 2026

A Rapid and Chemical-free Hemoglobin Assay with Photothermal Angular Light Scattering
Published on: December 7, 2016
BiliChek transcutaneous bilirubin meter overestimates serum bilirubin as measured by the Doumas reference method
Brad S Karon1, Andrea C Wickremasinghe, Stanley F Lo
1Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN 55905, USA. Karon.bradley@mayo.edu
Objectives:
To determine the relationship between BiliChek TcB (Respironics, Marietta GA) and Doumas reference serum or plasma total bilirubin (TSB).
Design And Methods:
Pooled samples with values assigned by the Doumas reference method were used to establish the relationship between a local laboratory and reference Doumas TSB. We then established the relationship between TcB and TSB in the 3 months before and after reassignment of calibrator setpoints undertaken to match the local laboratory to Doumas reference bilirubin values.
Results:
Before calibrator setpoint reassignment TSB as measured in our laboratory overestimated Doumas reference bilirubin. After calibrator adjustment laboratory TSB was within 1.7-6.8 micromol/L (0.1-0.4 mg/dL) of Doumas reference values. Mean bias between BiliChek TcB and TSB was 42.8+/-22.2 micromol/L (2.5+/-1.3mg/dL) (n=94) before and 49.6+/-22.2 micromol/L (2.9+/-1.3mg/dL) (n=115) after calibration adjustment.
Conclusions:
BiliChek TcB significantly overestimates TSB as measured by the Doumas reference method.
Related Concept Videos
Jaundice
Coulometry: Overview
Effect of Hepatic Disease on Pharmacokinetics: Pathophysiologic Assessment and Liver Function Test