Related Experiment Video
Updated: Jun 13, 2026

Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
Effect of bone quality on the artificial temporomandibular joint condylar prosthesis
Jui-Ting Hsu1, Heng-Li Huang, Ming-Gene Tu
1School of Dentistry, College of Medicine, China Medical University, Taichung, Taiwan. jthsu@mail.cmu.edu.tw
Objective:
This study evaluated the effect of bone quality on the stability of and stress in the artificial temporomandibular joint (TMJ) condylar prosthesis as well as on the stress and strain distributions in the bone.
Study Design:
Three-dimensional finite element models of the mandible and an artificial TMJ condylar prosthesis using 3 screws were established. Four types of bone were modeled by varying the elastic modulus of cancellous bone.
Results:
Increasing the bone quality from type IV to type I reduced the maximum relative micromotion between implant and bone, maximum von Mises stresses in the implant and surrounding cortical bone, and maximum von Mises strain in the surrounding cancellous bone by 17.5%, 10.5%, 12.0%, and 39.4%, respectively.
Conclusion:
The bone quality of the mandible has only a minor effect on the stability of and stress in an artificial TMJ implant and cortical bone, but it greatly affects the microstrain in cancellous bone.