[Thromboelastography/-metry and external quality control. Results of a pilot study]
A Dick1, M Schwaiger, C Jámbor
1Arbeitsgruppe Perioperative Hämostase, Abteilung für Transfusionsmedizin und Hämostaseologie, Klinik für Anästhesiologie, Klinikum der Universität München, 81377 München. andrea.dick@med.uni-muenchen.de
Hamostaseologie
|May 11, 2010
Summary
External quality assessment for Thromboelastography/Thromboelastometry (TEG/ROTEM) shows promising results. Lyophilized plasma samples enable reliable control of ROTEM results, ensuring accuracy in acute bleeding management.
Area of Science:
- Clinical pathology
- Hemostasis and thrombosis
- Point-of-care testing
Context:
- Thromboelastography/Thromboelastometry (TEG/ROTEM) is crucial for managing acute bleeding in perioperative and intensive care.
- A standardized external quality control for TEG/ROTEM results has been lacking.
- This study addresses the need for quality management in TEG/ROTEM analysis.
Purpose:
- To evaluate the feasibility of external quality assessment for Thromboelastometry (ROTEM) results.
- To assess the reproducibility of ROTEM parameters using lyophilized plasma samples.
- To establish a framework for quality control in near-patient hemostasis testing.
Summary:
- A pilot survey in 2008-2009 integrated into External Quality Assessment Schemes (EQAS) assessed ROTEM results.
- Lyophilized plasma samples were used, demonstrating good reproducibility for clot firmness (A20) and alpha-angle.
- Higher variability was observed for Clotting Time (CT) and Clot Formation Time (CFT), particularly with low fibrinogen levels.
Impact:
- This study demonstrates that ROTEM results can be reliably controlled through external quality schemes.
- Regular participation in external quality assurance can validate TEG/ROTEM as a beneficial technology in emergency settings.
- Improved quality control will enhance the reliability of TEG/ROTEM for acute bleeding management.