Double red cell concentrates -in vitro quality after delayed refrigeration
1Component Development Laboratory, NHS Blood and Transplant, Brentwood, UK. stephen.thomas@nhsbt.nhs.uk
Transfusion Medicine (Oxford, England)
|May 12, 2010
Summary
Automated collection of red cell concentrates (RCC) is feasible using the Haemonetics Cymbal device. Collected RCC can be stored at ambient temperature for up to 8 hours before refrigeration without compromising quality.
Area of Science:
- Transfusion Medicine
- Blood Component Collection Technology
Background:
- Automated collection of red cell concentrates (RCC) offers advantages for donors, blood services, and recipients.
- Remote collection sites can produce finished blood components.
- Limited data exist on the stability of collected RCC at ambient temperatures before refrigerated storage.
Purpose of the Study:
- To evaluate the quality of leucodepleted RCC collected using the Haemonetics Cymbal device.
- To determine the maximum permissible storage duration of collected RCC at ambient temperature before 4°C storage.
Main Methods:
- The Haemonetics Cymbal device collected RCC using citrate, phosphate, and dextrose (CPD-50) anticoagulant.
- Two leucodepleted RCC in saline, adenine, glucose, and mannitol (SAGM) additive solution were collected per procedure.
- One RCC from each pair was stored at ambient temperature for 6 hours, the other for 8 hours, before 42-day storage at 2-6°C.
- In vitro quality assessments included haemolysis, ATP, 2,3 DPG, and supernatant potassium levels.
Main Results:
- All collected RCC met UK and European standards for volume, haematocrit, and haemoglobin.
- Haemolysis levels at Day 42 were within specification and not significantly different between 6-hour and 8-hour warm storage groups.
- ATP, 2,3 DPG, and supernatant potassium levels were comparable between warm storage groups and reference data.
- Haemolysis levels tended to be higher than reference data for whole blood-derived RCC.
Conclusions:
- The Haemonetics Cymbal device can successfully collect two RCC in SAGM.
- In vitro data suggest that RCC collected via this method can be stored without refrigeration for up to 8 hours post-collection before 4°C storage.