Objective analysis of toric intraocular lens rotation and centration
James S Wolffsohn1, Phillip J Buckhurst
1Aston University, School of Life and Health Sciences, Ophthalmic Research Group, Birmingham, United Kingdom. j.s.w.wolffsohn@aston.ac.uk
Purpose:
To assess the repeatability of an objective image-analysis technique to determine intraocular lens (IOL) rotation and centration.
Setting:
Six European ophthalmology clinics.
Methods:
Eyes with an Akreos AO aspheric IOL with orientation marks were imaged. A masked observer rated image quality. The axis of rotation was determined from a line bisecting the IOL orientation marks. This was normalized for rotation of the eye between visits using the axis bisecting 2 consistent conjunctival vessels or iris features. The center of ovals overlaid to circumscribe the IOL optic edge and the pupil or limbus were compared to determine IOL centration. Intrasession repeatability was assessed in a subgroup of eyes and the variability of repeated analysis examined.
Results:
The SD of intrasession repeatability was +/-0.79 degrees for IOL rotational stability, +/-0.10 mm for horizontal centration, and +/-0.10 mm for vertical centration. The SD of repeated-analysis variability of the same image was +/-0.70 degrees, +/-0.20 mm, and +/-0.31 mm, respectively. The mean eye rotation (absolute) between visits was 2.23 degrees +/- 1.84 (SD) (10% >5 degrees rotation) using 1 set of consistent conjunctival vessels or iris features and 2.03 +/- 1.66 degrees (7%>5 degrees rotation) using the mean of 2 sets (P = .13). Poorer image quality resulted in larger apparent absolute IOL rotation (r = -0.45, P<.001).
Conclusions:
Objective analysis of digital retroillumination images allowed sensitive assessment of IOL rotation and centration stability. Eye rotation between images can lead to significant errors if not taken into account. Image quality is important to analysis accuracy.
Related Concept Videos
Torque Free Motion
Relative Motion Analysis using Rotating Axes
However, to express the relative position of point B relative to point A, an additional frame of reference, denoted as x'y', is necessary. This additional frame not only translates but also rotates relative to the fixed frame, making it instrumental in...
Properties of Enantiomers and Optical Activity
Rotation of Asymmetric Top
The relationship between the angular momentum of any rigid body and its angular velocity, both of which are vectors, involves the moment of inertia. The moment of inertia is a scalar quantity only for spherically symmetric...
Relative Motion Analysis using Rotating Axes-Problem Solving
Here, in order to determine the magnitude of velocity and acceleration for point...
Focusing of Light in the Eye