Investigation of optimal display size for detecting ground-glass opacity on high resolution computed tomography using
Michihiro Yamaguchi1, Hideki Fujita, Yuichi Bessho
1Department of Radiology, Osaka Prefectural Medical Center for Respiratory and Allergic Disease, Osaka Prefectural Hospital Organization, and Department of Medical Physics and Engineering, Graduate School of Medicine, Osaka University, 3-7-1 Habikino, Habikino City, Osaka 583-8588, Japan. yamaguchim@opho.jp
Abstract:
The purpose of this study was to clarify the relationship between display sizes of high resolution computed tomography (HRCT) images for detecting ground-glass opacity (GGO) and observer performance using a digital contrast-detail (d-CD) phantom. A structure of the d-CD phantom was determined on the basis of the actual images of GGOs and background noises of 22 patients who were diagnosed as GGO by chest HRCT. The d-CD phantom has a 512×512 matrix in size and has total of 100 holes: the diameter of these holes increases stepwise from 2 to 20 pixels with 2 pixels interval in a vertical direction and the CT value varies stepwise from 2 to 200 HU in a horizontal direction. The observer performance study was carried out for three different display sizes (30 cm×30 cm as an enlarged size, 13 cm×13 cm as an original size, and 7 cm×7 cm as a reduced size) using a 2-megapixels LCD monitor, and it was analyzed using Friedman and Wilcoxon statistical tests. As a result, the observer performance for the original display and the reduced display sizes was superior to that for the enlarged size (P=0.006 and 0.037 for the original display and the reduced display sizes, respectively), whereas there was no significant difference between the original display and reduced display sizes (P=0.77). The d-CD phantom enables a short-term evaluation of observer performance and is useful in analyzing relationship between display size and observer performance.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography