Inadequate gas supply to patients with an adjustable pressure-limiting valve in the fully opened position
Go Hirabayashi1, Hiroyuki Uchino, Yukihiko Ogihara
1Department of Anesthesiology, Hachioji Medical Center, Tokyo Medical University, 1163 Tatemachi, Hachioji, Tokyo, 193-0998, Japan. goh@tokyo-med.ac.jp
Abstract:
With spontaneous ventilation, sufficient exhaust resistance at the adjustable pressure-limiting (APL) valve when fully opened ensures that the reservoir bag fills and adequately supplies gas to patients. A lack of exhaust resistance with the APL valve fully open caused inadequate gas supply to patients with four types of anesthesia machines: SA2 (Dräger), Excel-210 SE (GE), Fabius (Dräger), and Cato (Dräger). Mechanically, the SA2 and Excel-210 SE APL valve systems, which are of the spring-loaded disc type positioned horizontally, cannot maintain sufficient exhaust resistance with the APL valve fully open. As for the Fabius and Cato, an exhaust valve independent of the APL valve should maintain sufficient exhaust resistance continuously. However, accumulated viscous substances on the thin diaphragm of the exhaust valve contributed to hindrance of diaphragm closure.
Related Concept Videos
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Mechanical Ventilation I: Indication and Settings
Pneumothorax II: Pathophysiology
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Atelectasis II: Pathophysiology
Pneumothorax-I
Pneumothorax can be even further classified as spontaneous, traumatic, and tension pneumothorax.