Effect of repeated applanation tonometry on the accuracy of intraocular pressure measurements
Dan D Gaton1, Miriam Ehrenberg, Moshe Lusky
1Department of Ophthalmology, Rabin Medical Center, Beilinson Campus, Petach Tikva, Israel. gaton@post.tau.ac.il
Purpose:
To evaluate the stability of successive applanation tonometry measurements in glaucomatous eyes.
Methods:
A prospective, comparative, randomized, and evaluator-masked study was conducted in a tertiary medical center. Sixty-seven patients with glaucoma attending our glaucoma clinic underwent four successive intraocular pressure (IOP) measurements with Goldmann-type applanation tonometry. Findings were compared with 70 patients scheduled for cataract surgery, similar in age and sex. The results were statistically analyzed with repeated measures analysis.
Results:
In the glaucoma group, the difference between the first and second IOP measurements was statistically significant (15.94 mmHg vs 14.9 mmHg, p < 0.0001), as was the decrease in IOP from the first measurement to each of the other successive measurements. No significant change in IOP was noted in the control group (mean, 13.7 mmHg).
Conclusions:
Patients with glaucoma show a decrease in IOP on repeated applanation tonometry measurements. By contrast, in individuals without glaucoma, no significant decrease in IOP on repeated applanation tonometry measurements was found.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Errors occurring during blood pressure monitoring
Several factors...
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Uncertainty in Measurement: Accuracy and Precision
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Accuracy and Precision