Epidural Intracranial Pressure Measurement in Rats Using a Fiber-optic Pressure Transducer
Published on: April 25, 2012
Pressure inside the neuroendoscope: correlation with epidural intracranial pressure during neuroendoscopic procedures
Lydia Salvador1, Ricard Valero, Jesús Carazo
1Department of Anesthesiology, Hospital Clínic de Barcelona, Barcelona, Spain.
Background:
During neuroendoscopic procedures, pressure inside the neuroendoscope (PIN) monitored through the irrigation channel correlates with the occurrence of postoperative complications. Our aim was to analyze the reliability of PIN measurement as a surrogate for intracranial pressure (ICP) by comparing PIN with simultaneously epidural ICP measurement as the standard.
Methods:
Seventeen consecutive patients undergoing neuroendoscopy were studied prospectively. Type and length of procedure and PIN and epidural ICP values during neuroendoscopy were recorded. Lin's concordance coefficient and Bland-Altman analysis of agreement were used to assess correspondence between the 2 systems.
Results:
A consistent relation between PIN and epidural ICP waveforms was observed during neuroendoscopic navigation. A strong Pearson correlation between PIN and epidural ICP data were found in 15 patients. Epidural ICP values were systematically higher than PIN values in 15 patients. Lin concordance coefficients showed moderate global agreement between the 2 methods, at 0.58 (95% confidence interval, 0.577-0.592). In 6 cases (35.2%) concordance was good according to this analysis, in 7 cases (41.2%) agreement was moderate/fair, and in 4 cases (23.5%) agreement was poor. The Bland-Altman analysis of patient data showed good agreement between the PIN and epidural ICP measurements for most patients, although discrepancies were greater at higher ICP values for 11 patients. Bland-Altman analysis of the complete dataset, after the normalization of individual's measurements, showed good overall agreement.
Conclusions:
PIN measurement seems useful for evaluating ICP changes related to neuroendoscopic procedures and seems to be more consistent than epidural ICP at high pressures.