A less sensitive detector does not necessarily result in a less sensitive method: fast quantification of 13 antiretroviral analytes in plasma with liquid chromatography coupled with tandem mass spectrometry
Related Experiment Video
Updated: Jun 12, 2026

Amplifying and Quantifying HIV-1 RNA in HIV Infected Individuals with Viral Loads Below the Limit of Detection by Standard Clinical Assays
Published on: September 26, 2011
Rob Ter Heine1, Hilde Rosing, Jos H Beijnen
1Department of Pharmacy and Pharmacology, Slotervaart Hospital, Amsterdam, The Netherlands. rob@terheine.nl
Background:
We previously developed a method for the simultaneous determination of the human immunodeficiency protease inhibitors: amprenavir, atazanavir, darunavir, indinavir, lopinavir, nelfinavir, ritonavir, saquinavir and tipranavir, the active nelfinavir metabolite M8 the non-nucleoside reverse transcriptase inhibitors efavirenz, nevirapine and etravirine and the internal standards dibenzepine, (13)C(6)-efavirenz, D5-saquinavir and D6-indinavir in plasma using liquid chromatography coupled with tandem mass spectrometry with a Sciex API3000 triple quadrupole mass spectrometer and an analytical run time of only 10 min. We report the transfer of this method from the API3000 to a supposedly less sensitive Sciex API365 mass spectrometer.
Methods:
We describe the steps that were undertaken to optimize the sensitivity and validation of the method that we transferred.
Results And Conclusions:
We showed that transfer of a method to a putative less sensitive detector did not necessarily result in a less sensitive assay, and this method can be applied in laboratories where older mass spectrometers are available. Ultimately, the performance of the method was validated. Accuracy and precision was within 87%-110% and <13%, respectively. No notable loss in selectivity was observed.
More Related Videos
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...