Repeatability of the Pedar-X in-shoe pressure measuring system
A K Ramanathan1, P Kiran, G P Arnold
1Institute of Motion Analysis & Research, Department of Orthopaedic and Trauma Surgery, TORT Centre, Ninewells Hospital & Medical School, University of Dundee, Dundee DD1 9SY, UK.
Background:
The Pedar-X is one of the newer versions of in-shoe pressure measuring devices and the current study aimed to assess the repeatability of this device.
Methods:
Twenty-seven healthy male volunteers were recruited and requested to walk on a 26-feet walkway wearing appropriate sized standardised off-the-shelf neutral running shoes (Donnay International). The Pedar-X insole was sandwiched between the foot and the shoe. Data were collected on two occasions, one week apart. Clinically relevant parameters studied were contact area, contact time in percentage roll over process, maximum force, pressure-time integral, force-time integral, peak pressure, mean force and mean area.
Results:
Repeatability was analysed using the coefficient of variation. Of the 160 parameters considered, 93.1% revealed a coefficient of variation value of less than 25. Heel and the metatarsal head areas were the most repeatable.
Conclusion:
The Pedar-X in-shoe pressure measuring system is repeatable and as such can be used as a valuable tool in the assessment of in-shoe plantar pressure distribution.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Assessment of blood pressure in brachial artery(two-step method)
Application of Pascal's Law
Hydraulic systems are used to operate automotive brakes, hydraulic jacks, and numerous other mechanical systems. We can derive a relationship between the forces in a simple hydraulic system by applying...
Measurement of Blood Pressure
Uncertainty in Measurement: Accuracy and Precision