Exogenous interferences with Jaffe creatinine assays: addition of sodium dodecyl sulfate to reagent eliminates
P Srisawasdi1, U Chaichanajarernkul, N Teerakanjana
1Division of Clinical Chemistry, Department of Pathology, Faculty of Medicine, Ramathibodi Hospital, Mahidol University, Bangkok, Thailand. srisawasdiP@yahoo.com
Background:
The study evaluated the impact of interferences on the analytical specificity of three commercial and commonly used creatinine methods (two Jaffe and one enzymatic).
Methods:
Manufacturer creatinine methods plus modified methods were tested with the following interferences: spiking serum with bilirubin, albumin, glucose, hemoglobin and lipid, and patient sera with maximum concentrations of bilirubin, 1,090 micromol/l and protein, 117.8 g/l.
Results:
Hemoglobin, 7.5 g/l and lipaemic with triglyceride concentration of 6.27 mmol/l, did not interfere with all assays. Glucose >33.3 mmol/l increased creatinine recovery for Dimension method. Samples spiked with bilirubin imparted a negative bias for Dimension and Architect methods but imparted a positive bias for Vitros assay. However, using patient sera, negative bias with bilirubin was found for all methods, from which Architect method gave the highest effect (R(2)=0.861), followed by Vitros (R(2)=0.239) and Dimension (R(2)=0.163). Protein provided the positive bias for all creatinine measurements that increased with increasing concentration (R(2) ranging from 0.104 to 0.182, P<0.0001). Addition of sodium dodecyl sulfate (SDS) in alkaline-picrate reagent reduced the effect of bilirubin and protein for kinetic Jaffe method. Although adding potassium ferricyanide was well effective for eliminating negative interference of bilirubin, it was prone to interference from protein.
Conclusions:
Endogenous interferences continue to plague creatinine accuracy measurement in both Jaffe and enzymatic methods, and consequentially the estimated glomerular filtration rate. The addition of SDS to the alkaline-pirate reagent was shown to be effective in reducing bilirubin and protein interferences.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Dialysis
EDTA: Indirect and Alkalimetric Titration
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...