Harrahill's technique: a simple screening test for intra-abdominal pressure measurement
J Otto1, M Binnebösel, K Junge
1Department of Surgery, University Hospital RWTH, Pauwelsstr. 30, 52074 Aachen, Germany. jeotto@ukaachen.de
Background:
Repair of giant incisional hernias may lead to an increase in intra-abdominal pressure (IAP) and, sometimes, to abdominal compartment syndrome. Measurement of IAP using Kron's technique (Kron et al. in Ann Surg 199:28-30, 1984) is currently accepted as the gold standard, whereas Harrahill has described a simple measurement setup using urinary drainage manometry (Harrahill in J Emerg Nurs 24:465-466, 1998). The aim of this clinical trail was to evaluate the correlation, reproducibility and effectiveness of this device.
Methods:
A prospective cohort study was performed in 43 patients undergoing elective standard abdominal intervention with laparotomy. These patients remain under surveillance in the intensive care unit and require a urinary catheter because of the operation. We performed comparative measurements of IAP using both Korn's (IVM) and Harrahill's (UDM) technique.
Results:
Evaluating the correlation between the IVM and UDM techniques, we measured median IAPs of 9.8 +/- 4.1 mmHg (2.9-19.9 mmHg) and 10.0 +/- 4.1 mmHg (min-max: 1.5-19.9 mmHg), respectively. Pearson's coefficient of correlation was r = 0.97. The average of difference between UDM and IVM was -0.2 +/- 0.9 mmHg with limits of agreement of -1.7 to 2.0 mmHg. Evaluating the reproducibility of Harrahill's technique, we found median IAPs of 10.4 +/- 2.1 mmHg (min-max: 2.9-19.1 mmHg) and 10.4 +/- 2.7 mmHg (3.7-19.9 mmHg), respectively, in 43 comparative measurements (Pearson's coefficient of correlation, r = 0.97. The average difference between both measurements was -0.1 +/- 1.1 mmHg with limits of agreement of -2.3 to 2.2 mmHg.
Conclusions:
We were able to demonstrate good correlation and high reproducibility of IAP measurement using Harrahill's technique compared to the gold standard Korn method. We consider this technique as a suitable method for quick and simple screening test for intra-abdominal hypertension, especially after repair of giant incisional hernias.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of the Abdomen III: Palpation