Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia
Published on: December 19, 2020
Multidetector CT of the lung: image quality with garnet-based detectors
Masahiro Yanagawa1, Noriyuki Tomiyama, Osamu Honda
1Department of Radiology, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita-city, Osaka 565-0871, Japan. m-yanagawa@radiol.med.osaka-u.ac.jp
Purpose:
To evaluate the image quality of both standard- and reduced-dose computed tomography (CT) by comparing multidetector CT with garnet-based detectors with multidetector CT with conventional detectors.
Materials And Methods:
The study was approved by the internal ethics review board. Informed consent was obtained. Eleven cadaveric lungs inflated and fixed by using the Heitzman method were scanned by using both CT with garnet-based detectors and CT with conventional detectors. Tube current was 400 mA for standard-dose and 10 mA for reduced-dose CT, and voltage was 120 kVp. Either normal scan mode with 984 views (conventional and garnet-based detectors) or high-resolution mode with 2496 views was used. Image quality at conventional-detector CT and garnet-based-detector CT in all modes was graded by two independent observers with a five-point scale. The evaluation items included normal lung structures, subjective visual noise, and abnormal CT findings. Quantitative image noise measurements were calculated by measuring the standard deviations in a circular region of interest on each selected image.
Results:
At standard-dose CT, image quality at CT with garnet-based detectors (high-resolution mode) was significantly improved (P < .001, Tukey-Kramer). However, there was no significant difference between quantitative image noise measurements (P > or = .24). At reduced-dose CT, only noise differed significantly, with both subjective visual noise and quantitative image noise measurements significantly greater at CT with garnet-based detectors (high-resolution mode) (P < or = .01). There was no significant difference in image quality except for noise between conventional-detector CT and garnet-based-detector CT (P > or = .06).
Conclusion:
The image quality of standard-dose garnet-based-detector CT (high-resolution) was significantly improved. Although highly reduced-dose garnet-based-detector CT (high-resolution mode) provided more image noise, overall image quality was not different between conventional-detector CT and garnet-based-detector CT.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT