Pulse wave velocity measurements are reproducible in multiple trained observers: a short report
M K Sigrist1, G Chiarelli, A Levin
1St. Paul's Hospital, Vancouver, B.C., Canada. msigrist@providencehealth.bc.ca
Background:
Arterial stiffness, measured by pulse wave velocity (PWV), is highly predictive of mortality in dialysis patients. As such, PWV is frequently used in clinical research studies and may have a role in clinical practice if shown to be suitably reliable. Measurement of PWV using the SphygmoCor system is known to be an observer-dependent technique. The aim of this study was to investigate the ability of 4 observers to acquire reproducible PWV and pulse wave analysis (PWA) measurements after a 6-week training period.
Methods:
Reproducibility of this technique was investigated using repeated measurements of the carotid-femoral PWV and PWA of the radial pulse by the 4 observers after a period of training. Both healthy volunteers and individuals with chronic kidney disease (CKD) were recruited for this study. Measurements were considered to have met quality control if 2 consecutive measurements were visually acceptable, within 1.5 m/s of each other and had a standard deviation of less than 10%. A fixed-effect analysis of variance was used to test the variation in measurements between the observers; the intraclass correlation coefficient (ICC) was used to assess the statistical agreement between the observers.
Results:
A total of 20 individuals volunteered for PWV and PWA measurements (13 with CKD and 7 without); the mean age was 58 years (range 24-83). The average PWV was 9.4 +/- 3.6 m/s. There was no significant difference shown between the 4 observers' measurements (p = 0.25). Further, there was good statistical agreement between the observers (ICC = 0.95).
Conclusions:
After a period of training it is possible for multiple observers to have reproducible measurements of PWV and PWA. Assurance of reproducibility is important when more than one individual is collecting data in a study, particularly when assessing changes over time.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring pulse
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: