Immunoglobulin interference in serum follicle-stimulating hormone assays: autoimmune and heterophilic antibody
Rachel Webster1, Michael Fahie-Wilson, Peter Barker
1Department of Clinical Biochemistry, Birmingham Children's Hospital, Steelhouse Lane, Birmingham, UK.
Abstract:
Interference in immunoassay caused by endogenous immunoglobulin is a cause of incorrect laboratory results that can drastically affect patient management. Two cases of immunoglobulin interference in serum follicle-stimulating hormone (FSH) assays are presented. These cases illustrate two common mechanisms for false-positive interference in two-site (sandwich) immunoassays. The first case describes a circulating autoimmune FSH immunoglobulin complex ('macro'-FSH), which has not been previously described for FSH, and the second a cross-linking antibody directed against the assay reagents. Immunoglobulin interference was detected and characterized using a combination of method comparison, immunosubtraction and size exclusion chromatography.