Purpose:
To compare the biomechanical properties of different numbers and locations of screws in a multidirectional volar fixed-angle plate in a distal radius osteotomy cadaver model.
Methods:
We created an extra-articular fracture in 16 pairs of fresh-frozen human cadaver radiuses. The 32 specimens were randomized into 4 groups. All fractures were fixated with a multidirectional volar fixed-angle plate. We tested 4 different screw-placement options in the distal fragment. The distal fragment was fixed with 4 locking screws in the distal row of the plate in group a, and with 4 locking screws alternately in the distal and proximal rows in group b. In group c, 3 locking screws were used in the proximal row; in group d, 7 locking screws were used, filling all screw holes in the distal and proximal rows of the plate. The proximal fragment was fixed with 3 screws. The specimens were loaded with 80 N under dorsal and volar bending and with 250 N axial loading. Finally, load to failure tests were performed.
Results:
Group d had the highest mean stiffness, 429 N/mm under axial compression, and was statistically significantly stiffer than the other groups. Group b had a mean stiffness of 208 N/mm, followed by group a, with 177 N/mm. Group c showed only a mean stiffness of 83 N/mm under axial compression. There were no statistically significant differences under dorsal and volar bending.
Conclusions:
In this model of distal radial fractures, there was a difference regarding the stiffness and the placement of screws in the distal rows of a volar fixed-angle plate. Inserting screws in all available holes in the distal fragment offered the highest stability. Using only the proximal row with 3 screws created an unstable situation. Based on these findings, we recommend placing at least 4 screws in the distal fragment and assigning at least 2 screws to the distal row of the multidirectional screw-holes.