Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Novel ultrasonic bone densitometry based on two longitudinal waves: significant correlation with pQCT measurement
1Department of Biosignal Pathophysiology, Kobe University Graduate School of Medicine, Kobe, Japan.
Unlabelled:
A reference database for trabecular bone density, cortical thickness, and elastic modulus of trabecular bone for a novel ultrasonic bone densitometry system (LD-100) based on two longitudinal waves (fast and slow) was determined over a wide age range in a normal Japanese population.
Introduction:
A novel ultrasonic bone densitometry system (LD-100 system) was applied to create a reference database for trabecular bone density (TBD), cortical thickness (CoTh), and elastic modulus of trabecular bone (EMTb) for this device over a wide age range in a normal Japanese population.
Methods:
In a comparative study between LD-100 and peripheral quantitative computed tomography (pQCT) systems, 52 individuals were examined by both systems at the same radius simultaneously. To create a reference database, a total of 2,380 healthy subjects (1,179 men, 1,201 women), ages 18-99 years, were examined using the LD-100 system.
Results:
Highly significant correlations between the LD-100 and pQCT systems were found in TBD (r = 0.877, p < 0.001) and CoTh (r = 0.723, p < 0.001). For the reference database, peak values of TBD, CoTh, and EMTb were observed at 30-34 years (255.09 mg/cm(3)), 20-24 years (5.23 mm), and 20-24 years (4.09 GPa) in men, and at 25-29 years (209.24 mg/cm(3)), 25-29 years (3.98 mm), and 20-24 years (3.33 GPa) in women, respectively. The TBD fell significantly (p < 0.05) beginning at 55-59 years in both sexes, with a relatively rapid decrease in women. The CoTh showed a significant decrease beginning at 40-44 years in men and 50-54 years in women. The EMTb showed a significant decrease beginning at 40-44 years in men and 55-59 years in women.
Conclusions:
The LD-100 system is a useful bone densitometry device and the database of age-related changes in TBD, CoTh, and EMTb established in this study will provide fundamental data for future studies related to bone status.