Related Experiment Video
Updated: Jun 12, 2026

The Nijmegen Hemostasis Assay: Simultaneous Fluorogenic Measurement of Thrombin and Plasmin Generation in a Single Well
Published on: February 27, 2026
Conventional chromogenic heparin assays are influenced by patient's endogenous plasma Antithrombin levels
1University of Alberta, Pediatrics, Edmonton, Canada. lesley.mitchell@albertahealthservices.ca
Objectives:
To determine, in vitro, the influence of plasma AT concentrations on heparin levels as measured by commercially available chromogenic kits.
Methods:
Purified AT was added to plasma that was immune depleted of AT at the following final concentrations: 0, 0.5, 1.0, 1.5, 2.0, 2.5 u/ml. Heparin was added to each of the AT concentrations at the following final concentrations: 0, 0.2, 0.4 u/ml. Heparin levels were then determined using 5 different Anti-Xa (n=4) or Anti-IIa (n=1) commercial kits. All kits provided purified AT to be added to the assay system.
Results:
When heparin was added to plasma, heparin concentrations measured were dependent on plasma concentrations of AT (p<0.001). When plasma concentrations of AT were less than 1.0 u/ml heparin concentrations were underestimated and when plasma concentrations of AT were greater than 1.0 u/ml actual heparin levels were over estimated. There was no significant difference in the findings between the various commercial Xa or IIa kits. In the absence of heparin, anticoagulant activity was detected when plasma AT concentrations were above 1.0 u/ml. There was a statistically significant correlation between plasma concentrations of AT and the amount of anticoagulant activity (p<0.001).
Conclusions:
Conventional chromogenic heparin assays are influenced by endogenous plasma AT levels.
Related Concept Videos
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Anticoagulant Drugs: Vitamin K Antagonists and Direct Oral Anticoagulants
Warfarin, a prominent vitamin K antagonist family member, exerts its effect by inhibiting the enzyme VKORC1 (vitamin K epoxide reductase complex 1). By hindering this enzyme, warfarin...
Venous Thrombosis III: Interprofessional Care
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...