Related Experiment Video
Updated: Jun 12, 2026

Dynamic Assessments of Coronary Flow Reserve after Myocardial Ischemia Reperfusion in Mice
Published on: August 25, 2023
Resting Pd/Pa measured with intracoronary pressure wire strongly predicts fractional flow reserve
Mamas A Mamas1, Simon Horner, Elise Welch
1Manchester Heart Centre, Manchester Royal Infirmary, Biomedical Research Centre, Manchester, United Kingdom.
Objective:
To investigate the relationship between resting distal coronary pressure to aortic pressure ratio (Pd/Pa) and fractional flow reserve (FFR) obtained during maximal hyperemia.
Background:
FFR is an invasive index of the functional severity of a coronary artery stenosis determined from coronary pressure measurements. It is generally believed that there is little correlation between resting Pd/Pa and FFR obtained during maximal hyperemia. We have therefore studied this relationship in a large cohort of patients who had undergone pressure- wire assessments.
Methods:
528 consecutive pressure-wire studies performed in 483 patients over a 2-year period were retrospectively analyzed.
Results:
A linear correlation between resting Pd/Pa and FFR post-pharmacological hyperemia was observed (rho = 0.74; p < 0.0001). When a FFR of < or = 0.75 (or < or = 0.80 as per FAME) was defined as positive, a resting Pd/Pa of < or = 0.85 (< or = 0.87) had a positive predictive value (PPV) of 95% (94.6%), while a resting Pd/Pa of > or = 0.93 (> or = 0.96) had a negative predictive value (NPV) of 95.7% (93%).
Conclusions:
We demonstrate a strong correlation between resting Pd/Pa and FFR. Resting values of Pd/Pa can be used to predict a positive FFR result with relatively high PPV and NPV. This may potentially obviate the need for adenosine infusion in a proportion of pressure-wire studies.