Related Experiment Videos

[A case experienced a difficulty in removing a catheter of IABP because of coagula formed in the balloon]

Y Terada1, K Yosida, T Saito

  • 1Department of Cardiovascular Surgery, Mitsuim Memorial Hospital.

Insights

A rare complication of intraaortic balloon pumping (IABP) occurred when a patient developed blood clots within the balloon, necessitating surgical removal. This complication likely resulted from a pinhole in the balloon allowing blood entry during procedures.

Area of Science:

  • Cardiology
  • Vascular Surgery
  • Medical Device Complications

Background:

  • Intra-aortic balloon pumping (IABP) is a crucial mechanical circulatory support device.
  • Complications, though rare, can arise from IABP use, impacting patient outcomes.
  • Severe atherosclerosis and ischemic heart disease are common indications for IABP.