Related Experiment Video
Updated: Jun 12, 2026

Measuring Spatially- and Directionally-varying Light Scattering from Biological Material
Published on: May 20, 2013
Calibration of a Shack-Hartmann wavefront sensor as an orthographic camera
J Vargas1, L González-Fernandez, J Antonio Quiroga
1Laboratorio de Instrumentación Espacial, Instituto Nacional de Técnica Aeroespacial,Carretera de Ajalvir Kilometer 4, 28850, Torrejón de Ardoz, Spain. jvargas@fis.ucm.es
Abstract:
We demonstrate a method to calibrate a Shack-Hartmann sensor as an orthographic camera. This calibration method permits us to obtain the distance, the rotation matrix between the microlens array and CCD imaging planes, and the projection matrix, which models the projection of the incoming rays to the CCD imaging plane. The proposed calibration method introduces a very compact matrix notation and allows wavefront reconstruction without an explicit centroid search between the reference and distorted spot diagrams. We show a set of simulations in code V that prove the effectiveness of the proposed method.