Examination of Rapid Dopamine Dynamics with Fast Scan Cyclic Voltammetry During Intra-oral Tastant Administration in Awake Rats
Published on: August 12, 2015
Rank estimation and the multivariate analysis of in vivo fast-scan cyclic voltammetric data
Richard B Keithley1, Regina M Carelli, R Mark Wightman
1Department of Chemistry, Neuroscience Center and Neurobiology Curriculum, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina 27599, USA.
Abstract:
Principal component regression has been used in the past to separate current contributions from different neuromodulators measured with in vivo fast-scan cyclic voltammetry. Traditionally, a percent cumulative variance approach has been used to determine the rank of the training set voltammetric matrix during model development; however, this approach suffers from several disadvantages including the use of arbitrary percentages and the requirement of extreme precision of training sets. Here, we propose that Malinowski's F-test, a method based on a statistical analysis of the variance contained within the training set, can be used to improve factor selection for the analysis of in vivo fast-scan cyclic voltammetric data. These two methods of rank estimation were compared at all steps in the calibration protocol including the number of principal components retained, overall noise levels, model validation as determined using a residual analysis procedure, and predicted concentration information. By analyzing 119 training sets from two different laboratories amassed over several years, we were able to gain insight into the heterogeneity of in vivo fast-scan cyclic voltammetric data and study how differences in factor selection propagate throughout the entire principal component regression analysis procedure. Visualizing cyclic voltammetric representations of the data contained in the retained and discarded principal components showed that using Malinowski's F-test for rank estimation of in vivo training sets allowed for noise to be more accurately removed. Malinowski's F-test also improved the robustness of our criterion for judging multivariate model validity, even though signal-to-noise ratios of the data varied. In addition, pH change was the majority noise carrier of in vivo training sets while dopamine prediction was more sensitive to noise.
More Related Videos
07:41Modeling Fast-scan Cyclic Voltammetry Data from Electrically Stimulated Dopamine Neurotransmission Data Using QNsim1.0
Published on: June 5, 2017
06:40Combined Infusion and Stimulation with Fast-Scan Cyclic Voltammetry (CIS-FSCV) to Assess Ventral Tegmental Area Receptor Regulation of Phasic Dopamine
Published on: April 23, 2020
Related Concept Videos
Voltammetric Techniques: Cyclic Voltammetry
Voltammetric Techniques: Pulse Voltammetry
Voltammetric Techniques: Linear-Scan (E vs Time)
Voltammetry: Factors Affecting Measurements
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...