Continuous non-invasive arterial pressure measurement: evaluation of CNAP device during vascular surgery
1Service d'anesthésie réanimation 1, hôpital Pellegrin, CHU de Bordeaux, place Amélie-Raba-Léon, 33076 Bordeaux cedex, France. matthieu.biais@chu-bordeaux.fr
Objective:
Standard non-invasive blood pressure (BP) monitoring is an intermittent, discontinuous procedure. Beat-to-beat BP monitoring requires invasive measurement via an arterial catheter and may be associated with serious complications. The Infinity CNAP SmartPod (Dräger Medical AG & Co. KG, Lübeck, Germany) has recently been proposed for non-invasive continuous beat-to-beat BP measurements. The present study was designed to compare BP obtained with the CNAP and with an invasive method in the operating room.
Study Design:
Prospective study.
Patients And Methods:
Twenty-five patients undergoing major vascular surgery were included. Systolic, mean and diastolic BP were monitored invasively (SAP, MAP and DAP respectively) and not invasively using the CNAP (CNAP-S, CNAP-M and CNAP-D respectively). Measurements were performed intraoperatively every minute during 1 hour.
Results:
One thousand and five hundred pairs of simultaneous CNAP and invasive BP measurements were obtained and 148 were eliminated. The range of BP measurements was 63-205 mmHg for SAP and 57-187 mmHg for CNAP-S, 38-143 mmHg for MAP and 43-142 mmHg for CNAP-M, 29-126 mmHg for DAP and 33-121 mmHg for CNAP-D. Bias and 95% limit of agreement between CNAP and invasive BP measurements were respectively 7.2 and -17.7 to 32.2 mmHg for SAP, -1.8 and -22.0 to 18.3 mmHg for MAP, and -7.5 and -27.3 to 12.4 mmHg for DAP. The percentage of CNAP measurements with a bias <10% with the arterial line was 69%, 86% and 91% for systolic, diastolic and mean pressures, respectively.
Conclusion:
Despite low accuracy for SAP and DAP measurements, CNAP system seems more accurate for MAP measurement in patients undergoing vascular surgery.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Aneurysm IV: Nursing Management
Peripheral Artery Disease V: Postoperative Nursing Management
Pre-Procedural Guidelines for Assessing Blood Pressure