Related Experiment Video
Updated: Jun 12, 2026

Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Cardiac CT: necessary for precise sizing for transcatheter aortic implantation
Carl J Schultz1, Adriaan D Moelker, Apostolos Tzikas
1Department of Cardiology, Erasmus MC Rotterdam, The Netherlands. c.schultz@erasmusmc.nl
Abstract:
Sizing for transcutaneous aortic valve replacement (TAVI) relies on non-invasive imaging. Incorrect sizing may result in adverse outcomes including paraprosthetic regurgitation, asymmetrical expansion which may impair prosthesis durability and, in the case of severe sizing errors, device embolisation or aortic root rupture. This review addresses the optimal approach for sizing. It is based on currently available data including the anatomical implications of using different imaging modalities, the steps of how to reproducibly measure the correct annulus diameter on MSCT, and current evidence for different sizing strategies.