Related Experiment Video
Updated: Jun 12, 2026

Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
Elimination technique for alkali metal ion adducts from an electrospray ionization process using an on-line ion
Kazuyoshi Nozaki1, Akira Tarui, Issey Osaka
1Department of Applied Chemistry, Faculty of Chemistry, Materials and Bioengineering, Kansai University, 3-3-35 Yamatecho, Suita, Osaka 564-8680, Japan.
Abstract:
The effects of an on-line ion suppressor device on alkali metal ion adduct formations of the model compound tacrolimus were investigated. The base peak ion in the positive ion ESI-MS spectrum of tacrolimus was a sodium ion adduct, [M+Na](+). On the other hand, an ammonium ion adduct, [M+NH(4)](+), was the base peak ion in the full-scan mass spectrum of tacrolimus with a cation-exchange suppressor resin, and both [M+Na](+) and [M+K](+) were eliminated. These results indicate that the combination of an on-line ion suppressor with ESI-MS is a simple and effective technique that eliminates undesirable alkali metal ion adduct formations in the positive-ion mode.
More Related Videos
11:04Ion Mobility-Mass Spectrometry Techniques for Determining the Structure and Mechanisms of Metal Ion Recognition and Redox Activity of Metal Binding Oligopeptides
Published on: September 7, 2019
10:41Ion Exchange Chromatography (IEX) Coupled to Multi-angle Light Scattering (MALS) for Protein Separation and Characterization
Published on: April 5, 2019
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Ion-Exchange Chromatography
Ion Exchange
Chemical Ionization (CI) Mass Spectrometry
Qualitative Analysis
For instance, group IV...
Atomic Absorption Spectroscopy: Interference