Related Experiment Video
Updated: Jun 12, 2026

Semi-Automated Phenotypic Analysis of Functional 3D Spheroid Cell Cultures
Published on: August 18, 2023
[Determination of pentostatin in culture broth using high performance liquid chromatography-mass spectrometry]
Peng Yang1, Yan Wang, Yanyan Liao
1Shanghai Institute of Pharmaceutical Industry, Shanghai 200040, China. chrisyang_1984@hotmail.com
Abstract:
A method for the determination of pentostatin in culture broth was established using high performance liquid chromatography-mass spectrometry (HPLC-MS). The chromatographic conditions used were as follows: the analytical column was Hypersil ODS2 (250 mm x 4.6 mm, 5 microm); the mobile phase was methanol/acetonitrile/10 mmol/L ammonium acetate (pH 7.6) (2.5/2. 5/95, v/v/v) with a flow rate of 1.0 mL/min; the detection wavelength was 280 nm; the injection volume was 10 microL; the temperature was 40 degrees C. The linear range of the method for pentostatin was 1.0 - 100 mg/L (r = 0.999 9). The method is characterized by simplicity, higher sensitivity and accuracy.