Related Experiment Video
Updated: Jun 12, 2026

Treating Low Back Pain in Failed Back Surgery Patients with Multicolumn-lead Spinal Cord Stimulation
Published on: June 26, 2018
Spontaneous lead breakage in implanted spinal cord stimulation systems
Tae Hun Kim1, Pyung Bok Lee, Hye Min Son
1Department of Anesthesiology and Pain Medicine, Seoul National University School of Medicine, Seoul, Korea.
Abstract:
Spinal cord stimulation (SCS) has become an established clinical option for treatment of refractory chronic pain. Current hardware and implantation techniques for SCS are already highly developed and continuously improving; however, equipment failures over the course of long-term treatment are still encountered in a relatively high proportion of the cases treated with it. Percutaneous SCS leads seem to be particularly prone to dislocation and insulation failures. We describe our experience of lead breakage in the inserted spinal cord stimulator to a complex regional pain syndrome patient who obtained satisfactory pain relief after the revision of SCS.
Related Concept Videos
Secondary Spinal Cord Injury llI: Pathophysiology
Spinal Cord Injury ll: Pathophysiology