Current uses of transfusion administration sets: a cause for concern?
M J Nightingale1, D R Norfolk, D J Pinchon
1NHS Blood and Transplant, Southampton, UK. mark.nightingale@nbs.nhs.uk
Modern infusion pumps pose risks with standard transfusion sets. Current standards (ISO 1135-4) are inadequate, necessitating updated guidelines for manufacturers and users to ensure patient safety during blood transfusions.
Area of Science:
- Medical Devices
- Biomedical Engineering
- Patient Safety
Background:
- Transfusion administration sets are used with infusion pumps for blood component delivery.
- Modern infusion pumps operate at negative pressures exceeding the capabilities of current transfusion set standards.
- Inadequate compatibility between infusion pumps and transfusion sets can compromise patient safety.
Purpose of the Study:
- Evaluate the adequacy of International Standard ISO 1135-4 for current transfusion set applications.
- Assess the suitability of UK-supplied infusion pumps for blood component administration.
- Identify necessary control measures for manufacturers and users to enhance patient safety.
Main Methods:
- Conducted a keyword literature search correlating transfusion parameters with adverse effects.
- Standardized units for occlusion pressure, flow rate, and hemolysis for comparison.
- Reviewed transfusion set instructions for use and surveyed UK infusion pump suppliers regarding their specifications and compatible sets.
Main Results:
- Clinically unacceptable hemolysis was previously shown above 40 kPa.
- Modern infusion pumps generate negative pressures up to 210 kPa, exceeding ISO 1135-4 testing capabilities (which omits negative pressure testing).
- Transfusion sets lacked validated information on flow/pressure restrictions and compatible blood components.
Conclusions:
- International Standard ISO 1135-4 requires updating to address modern infusion pump performance, particularly negative pressure.
- Recommendations are provided for manufacturers on pressure limitations and blood component compatibility.
- Users need guidance on selecting, configuring, and validating infusion pumps and associated disposables for safe blood transfusion.
Related Concept Videos
Blood Transfusion and Agglutination
History
The history of blood transfusion dates back to the 17th century, when early attempts were made in animals. In 1818 James Blundell, a British doctor, performed the first successful human blood transfusion. Later in 1900, Karl...
Blood Transfusion
Blood Transfusion Overview
A blood transfusion is a medical procedure used to replace blood lost due to injury, surgery, or to treat conditions such as anemia or cancer. During a transfusion, donor blood is...
Tissue Transplantation
The Biology of Tissue Transplantation
The biology of tissue transplantation hinges on the Major Histocompatibility Complex (MHC) molecules. These molecules...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Errors occurring during blood pressure monitoring
Several factors...