X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
The characteristics of dental X-ray fluoroscopic equipment 'DreamRay 60F'
Gwi-Soon Shin1, Bo-Ram Lee, Kyung-Hwan Chang
1Medical Physics Laboratory, Department of Radiological Technology, College of Health Science, Korea University, Seoul, Republic of Korea.
Abstract:
This study examined the characteristics of the dental X-ray fluoroscopic equipment, 'DreamRay 60F', which was recently developed in Korea. The output linearity, output reproducibility, half-value layer (HVL), leakage radiation and scattered radiation were measured using an ionisation chamber. The surface dose equivalent rate and estimated dose equivalent of the operator were also calculated. The output linearity was 0.0015-0.0175 and the coefficient of variation for the output reproducibility was 0.0013-0.0074. The experimental HVL was 2.1 mm Al, and the leakage dose rate at 100 cm from the X-ray focus ranged from 2.70 to 19.66 microGy h (-1) depending on the direction. The scattered radiation doses differed significantly (1.7-16.8 times) depending on the distance and direction. If an operator is exposed for 10 min per procedure, 5 procedures a day at 5 days a week, he/she sitting at a 90 degrees direction will receive an annual dose equivalent of 13.0 mSv (at 30 cm) and 63.7 mSv (at 50 cm) in the trunk and face surface, respectively.
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...