A simple pacing method to diagnose postero-anterior (clockwise) cavo-tricuspid isthmus block after radiofrequency

Agustín Pastor1, Ambrosio Núñez, Gonzalo Guzzo

  • 1Cardiology Service of Hospital Universitario de Getafe, Ctra de Toledo, Getafe, Madrid, Spain.

Insights

A new pacing method reliably confirms clockwise cavo-tricuspid isthmus (CTI) block after atrial flutter ablation, significantly reducing flutter recurrence. This technique improves diagnosis when other methods are unclear.

Area of Science:

  • Cardiology
  • Electrophysiology
  • Medical Devices

Background:

  • Bidirectional block of the cavo-tricuspid isthmus (CTI) is crucial for typical atrial flutter ablation success.
  • Evaluating clockwise CTI block can be challenging, potentially impacting ablation outcomes.

Purpose of the Study:

  • To assess pacing at the septal edge of the ablation line as an indicator of clockwise CTI block.
  • To determine if this method predicts atrial flutter recurrence.

Main Methods:

  • Ninety-four patients undergoing flutter ablation were studied.
  • Clockwise CTI block was evaluated using differential septal pacing and septal CTI pacing.
  • Electrogram (EGM) analysis was performed during pacing maneuvers.

Main Results:

  • Bidirectional CTI block was achieved in 78% of patients.
  • Persistent conduction was observed in 13%, associated with a 48% flutter recurrence rate.
  • Septal CTI pacing reliably diagnosed clockwise CTI block when differential septal pacing EGMs were difficult to interpret (36% of cases).

Conclusions:

  • Septal CTI pacing is a reliable and simple method for diagnosing clockwise CTI block.
  • This diagnostic approach is linked to a low rate of atrial flutter recurrence.
Abstract