Signal to cut-off (S/CO) ratio and detection of HCV genotype 1 by real-time PCR one-step method: is there any direct
Guilherme Albertoni1, C P Arnoni, P R B Araújo
1Associação Beneficente de Coleta de Sangue, Av. Jandira 1260, São Paulo, SP, Brazil. albertonig@nefro.epm.br
Background:
Polymerase chain reaction (PCR) methods play an essential role in providing data related to diagnosis, monitoring and treatment of hepatitis C virus (HCV) infection. EIA results are reported as ''reactive'' or ''non reactive'' and EIA S/CO ratio may also be reported as ''high'' or ''low.'' This study aimed to evaluate the performance of a real-time RT-PCR and assess whether there is relationship between S/CO and PCR results.
Study Design And Methods:
Sera from blood donors were analyzed by Enzyme-Linked Immunosorbent Assay (ELISA) and RT-PCR assay to detect HCV infection.
Results:
The RT-PCR assay to genotypes 1a/b showed an acceptable linear response in serial dilutions. The samples were divided into two groups based on their serological results: group A--S/CO ratio < 3 (60 samples) and group B--S/CO ratio > 3 (41 samples). Viral loads were confirmed positive in group B samples in 90%, and in group A samples were confirmed positive in only 13% by RT-PCR.
Conclusion:
The methodology used was able to detect the presence of RNA-HCV genotype I in 90% of the samples serologically positive in group B. All negative samples were sent to search for other genotypes of HCV (genotypes 2-6) and were confirmed as negative. These data suggests that these negative samples may have HCV RNA viral load below the detection limit of our test (310 IU/ mL), or a false positive result in serological test, or spontaneous viral clearance occurred.