Phase shifting interferometry in the presence of vibration: a new algorithm and system
Abstract:
A modification to the usual phase shifting interferometry algorithm permits measurements to be taken fast enough to essentially freeze out vibrations. Only two interferograms are time critical in this 2 + 1 algorithm; the third is null. The implemented system acquires the two time critical interferograms with a 1-millisecond separation on either side of the interline transfer of a standard CCD video camera, resulting in a reduction in sensitivity to vibration of 1-2 orders of magnitude. The required phase shift is achieved via frequency shifting. Laboratory tests comparing this system with a commercial phase shifting package reveal comparable rms errors when vibrations are low; as expected from an analysis of potential phase errors. However, the 2 + 1 system also succeeded when vibrations were large enough to wash out video rate fringes.
Related Concept Videos
IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations
Interference and Diffraction
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...