X-ray fluorescence (XRF) set-up with a low power X-ray tube
Sheenu Gupta1, Kanan Deep, Lalita Jain
1Nuclear Science Laboratories, Physics Department, Punjabi University, Patiala 147 002, India.
Abstract:
The X-ray fluorescence set-up with a 100 W X-ray tube comprises a computer controlled system developed for remote operation and monitoring of tube and an adjustable stable 3D arrangement to procure variable excitation energies with low scattered background. The system was tested at different filament currents/anode voltages. The MDL of the set-up at 0.05-1.00 mA/4-12 kV is found approximately (1-100)ppm for K and L excitations and approximately (200-700)ppm for M excitations of elements and improves with filament current and anode voltage. Moreover, L measurements for Sm and Eu at five K X-ray energies of elements(Z=29-40) and analytical determination in some synthetic samples were undertaken.