Related Experiment Video
Updated: Jun 12, 2026

In Vivo Morphometric Analysis of Human Cranial Nerves Using Magnetic Resonance Imaging in Menière's Disease Ears and Normal Hearing Ears
Published on: February 21, 2018
Cartilage morphology at 3.0T: assessment of three-dimensional magnetic resonance imaging techniques
Christina A Chen1, Richard Kijowski, Lauren M Shapiro
1Department of Radiology, Stanford University, Stanford, California, USA.
Purpose:
To compare six new three-dimensional (3D) magnetic resonance (MR) methods for evaluating knee cartilage at 3.0T.
Materials And Methods:
We compared: fast-spin-echo cube (FSE-Cube), vastly undersampled isotropic projection reconstruction balanced steady-state free precession (VIPR-bSSFP), iterative decomposition of water and fat with echo asymmetry and least-squares estimation combined with spoiled gradient echo (IDEAL-SPGR) and gradient echo (IDEAL-GRASS), multiecho in steady-state acquisition (MENSA), and coherent oscillatory state acquisition for manipulation of image contrast (COSMIC). Five-minute sequences were performed twice on 10 healthy volunteers and once on five osteoarthritis (OA) patients. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured from the volunteers. Images of the five volunteers and the five OA patients were ranked on tissue contrast, articular surface clarity, reformat quality, and lesion conspicuity. FSE-Cube and VIPR-bSSFP were compared to IDEAL-SPGR for cartilage volume measurements.
Results:
FSE-Cube had top rankings for lesion conspicuity, overall SNR, and CNR (P < 0.02). VIPR-bSSFP had top rankings in tissue contrast and articular surface clarity. VIPR and FSE-Cube tied for best in reformatting ability. FSE-Cube and VIPR-bSSFP compared favorably to IDEAL-SPGR in accuracy and precision of cartilage volume measurements.
Conclusion:
FSE-Cube and VIPR-bSSFP produce high image quality with accurate volume measurement of knee cartilage.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...