Related Experiment Video
Updated: Jun 12, 2026

In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Comparison of postprocessing techniques for the detection of perfusion defects by cardiac computed tomography in
Ian S Rogers1, Ricardo C Cury, Ron Blankstein
1Department of Radiology and Division of Cardiology, Massachusetts General Hospital, Boston, MA 02114-2750, USA.
Background:
Despite rapid advances in cardiac computed tomography (CT), a strategy for optimal visualization of perfusion abnormalities on CT has yet to be validated.
Objective:
We evaluated the performance of several postprocessing techniques of source data sets to detect and characterize perfusion defects in acute myocardial infarctions with cardiac CT.
Methods:
Twenty-one subjects (18 men; 60 +/- 13 years) that were successfully treated with percutaneous coronary intervention for ST-segment myocardial infarction underwent 64-slice cardiac CT and 1.5 Tesla cardiac magnetic resonance imaging (MRI) scans after revascularization. Delayed enhancement MR images were analyzed to identify the location of infarcted myocardium. Contiguous short-axis images of the left ventricular myocardium were created from the CT source images with 0.75-mm multiplanar reconstruction (MPR), 5-mm MPR, 5-mm maximal intensity projection (MIP), and 5-mm minimum intensity projection (MinIP) techniques. Segments already confirmed to contain infarction by MRI were then evaluated qualitatively and quantitatively with CT.
Results:
Overall, 143 myocardial segments were analyzed. On qualitative analysis, the MinIP and thick MPR techniques had greater visibility and definition than the thin MPR and MIP techniques (P < 0.001). On quantitative analysis, the absolute difference in Hounsfield unit attenuation between normal and infarcted segments was significantly greater for the MinIP (65.4 Hounsfield unit [HU]) and thin MPR (61.2 HU) techniques. However, the relative difference in Hounsfield unit attenuation was significantly greatest for the MinIP technique alone (95%; P < 0.001). Contrast to noise was greatest for the MinIP (4.2) and thick MPR (4.1) techniques (P < 0.001).
Conclusion:
The results of our current investigation found that MinIP and thick MPR detected infarcted myocardium with greater visibility and definition than MIP and thin MPR.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...