Forearm and upper-arm oscillometric blood pressure comparison in acutely ill adults
Kathleen Schell1, Kate Morse, Julie K Waterhouse
1University of Delaware, Newark, USA. kaschell@udel.edu
Abstract:
When patients' upper arms are not accessible and/or when cuffs do not fit large upper arms, the forearm site is often used for blood pressure (BP) measurement. The purpose of this study is to compare forearm and upper-arm BPs in 70 acutely ill adults, admitted to a community hospital's 14-bed ICU. Using Philips oscillometric monitors, three repeated measures of forearm and upper-arm BPs are obtained with head of bed flat and with head of bed elevated at 30 degrees. Arms are resting on the bed. Paired t tests show statistically significant differences in systolic BPs, diastolic BPs, and mean arterial pressures in the supine and head-elevated positions. Bland-Altman analyses indicate that forearm and upper-arm oscillometric BPs are not interchangeable in acutely ill adults.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: