Employing the Forced Oscillation Technique for the Assessment of Respiratory Mechanics in Adults
Published on: February 9, 2022
Intra- and intersession repeatability of a double-pass instrument
Meritxell Vilaseca1, Elvira Peris, Jaume Pujol
1Centre for Sensors, Instruments and Systems Development (CD6), Universitat Politécnica de Catalunya (UPC), Terrassa, Barcelona, Spain. mvilasec@oo.upc.edu
The Optical Quality Analysis System (OQAS) demonstrates good repeatability for measuring optical quality parameters, even with eye realignment. This ensures reliable clinical assessments of visual function.
Area of Science:
- Ophthalmology
- Optical Engineering
- Biomedical Instrumentation
Background:
- The Optical Quality Analysis System (OQAS) utilizes the double-pass technique to assess ocular optical quality.
- Accurate and repeatable measurements are crucial for clinical evaluation of visual function.
Purpose of the Study:
- To evaluate the intra- and intersession repeatability of optical quality parameters obtained using the OQAS.
- To determine the impact of eye realignment on measurement variability.
Main Methods:
- OQAS measurements were performed on 20 healthy eyes (best spectacle-corrected visual acuity ≥ 20/20).
- Examinations were conducted across three separate sessions with 10-minute intervals, including eye realignment at each session start.
- Within each session, three consecutive measurements were taken without realignment.
- Analyzed parameters included MTFcutoff, Strehl ratio, OQAS values (OV) at various contrasts, and objective scatter index (OSI).
Main Results:
- Mean coefficients of repeatability for intra- and intersession measurements were consistent across all analyzed parameters (MTFcutoff, Strehl ratio, OV at 100%, 20%, 9%, and OSI).
- Bland and Altman analysis showed confidence limits indicating good agreement for both intra- and intersession repeatability.
- No significant additional variability was introduced by eye realignment between sessions.
Conclusions:
- The OQAS is a reliable clinical instrument for assessing ocular optical quality.
- The system exhibits good intra- and intersession repeatability.
- Eye realignment does not significantly affect the variability of OQAS measurements.
More Related Videos
Related Concept Videos
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Bandpass Sampling
A bandpass signal has a spectrum with a lower frequency limit, denoted as ω1, and an upper frequency limit, denoted as ω2. The spectrum...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Reconstruction of Signal using Interpolation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.