Standard photoacoustic spectrometer: model and validation using O2 A-band spectra
K A Gillis1, D K Havey, J T Hodges
1Process Measurements Division, National Institute of Standards and Technology, Gaithersburg, Maryland 20899, USA. keith.gillis@nist.gov
Abstract:
We model and measure the absolute response of an intensity-modulated photoacoustic spectrometer comprising a 10 cm long resonator and having a Q-factor of approximately 30. We present a detailed theoretical analysis of the system and predict its response as a function of gas properties, resonance frequency, and sample energy transfer relaxation rates. We use a low-power continuous wave laser to probe O(2) A-band absorption transitions using atmospheric, humidified air as the sample gas to calibrate the system. This approach provides a convenient and well-characterized method for calibrating the absolute response of the system provided that water-vapor-mediated relaxation effects are properly taken into account. We show that for photoacoustic spectroscopy (PAS) of the O(2) A-band, the maximum conversion efficiency of absorbed photon energy to acoustic energy is approximately 40% and is limited by finite collision-induced relaxation rates between the two lowest-lying excited electronic states of O(2). PAS also shows great potential for high-resolution line shape measurements: calculated and experimental values for the PAS system response differ by about 1%.
Related Concept Videos
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
UV–Vis Spectrometers
UV–Vis Spectroscopy: Woodward–Fieser Rules
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...