The frequency of electrocardiographic errors due to electrode cable switches: a before and after study
Thomas Thaler1, Volker Tempelmann, Marco Maggiorini
1Department of Cardiology, Canton Hospital Sankt Gallen, Rorschacher Strasse 95, Sankt Gallen, Switzerland.
Introduction:
Electrocardiographic (ECG) errors due to electrode cable reversal confuse physicians and provoke unnecessary diagnostic tests. They occur in approximately 4% of ECGs performed in the intensive care unit (ICU). We aimed to investigate whether this frequency could be reduced by an appropriate intervention.
Methods:
All ECGs from consecutive patients were collected at ICU discharge and analyzed by the investigators. Before collecting a second set of ECGs, we educated our ICU staff and performed technical improvements on the electrocardiographs (system approach). Electrocardiographic errors were identified applying previously published morphologic criteria.
Results:
We collected and analyzed 1123 ECGs from 416 patients. Nine hundred ten ECGs (81%) were recorded in the ICU; and the frequencies of electrode cable misplacements before and after the intervention were 4.8% and 1.2%, respectively (P = .002).
Conclusions:
Using a system approach, we were able to significantly reduce the frequency of ECG errors due to electrode cable switches by 75%.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Dysrhythmias V: Evaluating Dysrhythmias
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...